APPLICATION:
Air-cooled Weather Conditions Meter. To determine the color fastness and aging
to sunlight, weather resistance, compound color fastness to
sunlight-perspiration of colored textile, leather, leatheroid, plastic, etc, by
setting various parameters of irradiance, temperature, humidity, spray to
simulate natural condition in lab. Be able to control irradiance real-time,
automatically inspect and compensate irradiance, temperature, humidity.
RELATED STANDARDS:
|
STANDARDS |
ISO
105-B02, B04,
B06 |
Note: this tester can conform to but not limit to all the standards above, for
more standards conformance, please contact us.
FEATURES:
l
Presetting various specifications, meet diverse range of standards, including
ISO, etc.
l
Light source:1500 watt air cooled xenon arc lamp;
l
Blue LCD
Panel, the operator simply selects the required test module and test standard;
easy
to use.
l
Rotate model: revolution, rotation.
l
Irradiation control,
24
steps adjustable,
manual control;
l
Manageable
size is most appropriate to testing fabric samples, and easy access to sample
holder;
l
Ultrasonic humidification system makes spray water even and low noise;
l
Chamber temperature control, irradiation temperature, heater raise the
temperature;
KEY SPECIFICATION:
l
Power
supply
Ac220V±5%
50Hz
20A
2500W
l
Irradiation Source:
1500W Air-cooled Long Xenon Arc Lamp
l
Average Lamp Life:
1000 hours
l
Rack Rotating Speed:
5r/min
l
Diameter of
Sample
Rack
350mm
l
Specimen Holder Exposure Area: 100×45mm(ISO)
l
Max
Exposure
Area
of
Specimen 450
l
BST
30℃~110℃±1℃
l
BPT
40℃~80℃±2℃
l
Humidity
Range 15%RH~95%±5%RH
l
Irradiance range
290nm~800nm
l
Instrument
Dimension
850×620×1750mm(L×W×H)
l
Instrument Weight 180kg
ENVIRONMENTAL REQUIREMENTS (AMBIENT):
Temperature
Range:
16-29℃
(60- 85℉)
Humidity
Range:
0-85% RH, non-condensing.
Altitude:
Up to 2000 meters (6560 feet)
Storage:
10-43℃
(50-110℉), 20-80% RH
Installation Category: II
Pollution Degree:
2
The room should be ventilated properly with minimal drafts, air pollution, or
dust. Do not install the instrument in a corrosion testing environment, or
adjacent to corrosion testing equipment it is not advisable to install the
instrument in such locations as a warehouse, boiler room, garage/factory are,
paint room, or where doors to the outside environment are frequently opened
causing changes in room temperature and humidity. Such locations may adversely
affect instrument operation, longevity and test integrity (repeatability).
RELATED KEY WORDS:
colour fastness to washing,colour measurement instrument,colourfastness to hot
pressing heating,colourfastness to perspiration,colourfastness to
rubbing,colourfastness to washing,Comb sorters,compatibility testing,Compressing
strength,compression strength,compression test equipment,compression test
machine,compression tester,compression testing,compression testing machine |