APPLICATION:
Water-cooled Weather Conditions Meter. To determine the color fastness and aging to sunlight, weather resistance, compound color fastness to sunlight-sweat of colored textile, leather, leatheroid, plastic, etc, by setting various parameters of irradiance, temperature, humidity, spray to simulate natural condition in lab. Be able to control irradiance real-time, automatically inspect and compensate irradiance, temperature, humidity.
RELATED STANDARDS:
|
STANDARDS |
ISO 105-B02/B04/B06, 4892-2, 11341 |
AATCC TM16, TM169 |
JIS L0843, GMW 3414 | Note: this tester can conform to but not limit to all the standards above, for more standards conformance, please contact us.
FEATURES:
● Presetting various specifications, or customized programs, meet diverse range of standards, including AATCC, ISO, BS, JIS etc;
● EviewTM touch screen, and build-in PLC (OMRONTM, Japan), real-time display the ongoing test’s graphic, be able to presetting irradiance, temperature, humidity, and plot dynamic curve of them.
● 4500W water-cooled xenon arc lamp system, accurately simulate sunlight spectrum. Irradiance auto- compensates system, control irradiance, temperature, and humidity timely through a closed loop.
● Irradiation control, light irradiance sensor monitor, digital setting, automatic compensating system, stepless irradiation regulation.
● Accurate chamber temperature control, irradiation temperature, and heater raise the temperature, air conditioning refrigeration system.
● Ultrasonic humidification system makes spray water even and low noise. Saturated steam humidifying, industrial refrigerating machine dehumidifying, and close loop control the humidity.
● Large amount of testing specimen, be able to hold as 3 times of specimen quantity as air-cooled machine in a test.
● Custom program capability and storage for additional user-defined tests.
● De-ionized water system to ensure purity of cooling water and spray water.
● Each specimen holder has time recorder respectively. Ease to test different specimen in one test.
● Multi-point inspection: malfunction alarm, self-diagnostic function, ease to maintenance.
● Print the test result chart by stylus printer directly. No need support of PC.
KEY SPECIFICATION:
● Power Supply: AC220V±5% 50Hz 9000W(Max)
● Irradiation Source: 4500W Water-cooled Long Xenon Arc Lamp
● Average Lamp Life: 1000 hours/1500 hours ()
● Rack Rotating Speed: 3r/min
● Rack Outside Radius: 224mm (Centered on the Xenon-arc Lamp)
● Specimen Holder Exposure Area: 100×45mm(ISO) 120×60mm(AATCC)
● Specimen Number: 29 pcs(ISO); 17pcs(AATCC)(Max)
● Exposure Area of Specimen ≥2800
● Chamber Temperature Range: 25℃~85℃±2℃
● BPT: 20℃~120℃±2℃
● BST: 20℃~130℃±1℃
● Humidity
Light Cycle: 10%RH~85%RH ±5% RH
Dark Cycle: 10%RH~100%RH ± 5% RH
● Irradiance range
Wide Bandwidth Measurement: (35~150)±1W/m2 @300~400nm
Narrow Bandwidth Measurement: (0.800~3.000)±0.02W/m2 @420nm
Or specified by the custom with 340nm, 300~800nm.
● Compressor Power: 1500W
● DI water Container: 5000mL
● Floor Space: 1050×3000mm (Min)
● Instrument Dimension: 1200×670×1800mm(, , L×W×H)
● Instrument Weight: 400kg
STRUCTURE INSIDE THE EXPOSURE CHAMBER:
① Water-Cooler Xenon-arc Lamp.
② Irradiation Degree Sensor
③ BST, BPT, Specimen holder, etc.
④ Specimen Rack
⑤ Temperature, Humidity Sensor
⑥ Spray nozzle (Front)
⑦ Spray nozzle (Back)
① Water-Cooled Xenon-arc Lamp and the Light Filter
②Specimen Rack
③Specimen holder for ISO and AATCC
④ Online Interface
Touch screen, be able to presetting irradiance, temperature, humidity, and plotting dynamic curve of them and control the instrument at real-time.
ENVIRONMENTAL REQUIREMENTS (AMBIENT):
● Temperature Range: 16-29℃ (60- 85oF)
● Humidity Range: 0-85% RH, non-condensing.
● Altitude: Up to 2000 meters (6560 feet)
● Storage: 10-43℃ (50-110oF), 20-80% RH
● Installation Category: II
● Pollution Degree: 2
The room should be ventilated properly with minimal drafts, air pollution, or dust. Do not install the instrument in a corrosion testing environment, or adjacent to corrosion testing equipment it is not advisable to install the instrument in such locations as a warehouse, boiler room, garage/factory are, paint room, or where doors to the outside environment are frequently opened causing changes in room temperature and humidity. Such locations may adversely affect instrument operation, longevity and test integrity (repeatability).
RELATED KEY WORDS:
chemicals for perspiration test,china equipment,China Infrared Sample Dyeing Machine,cloth test,cloth tester,cloth testing,Cloth Testing Instrument,color assessment,color assessment cabinet,COLOR CHEMICAL TESTING MACHINE,color evaluation,color fast,color fastness,color fastness in,color fastness of |